Advanced Programming 7-15
September 2007
GAS VALUE EXP FILTER
Option (cont.)
The Set Window Threshold option (cont.):
You may now do one of the following:
• Press [N] as many times as necessary to select another Gas Value
Exp Filter menu option. Then, press
[Y] and proceed to the
appropriate section of this chapter for instructions.
• Press [N] until you reach the Done? option of the Gas Value Exp
Filter menu, then press
[Y] to return to the Calibrate System menu.
The Set Smooth Threshold option:
The smooth threshold, which is calculated from the oxygen span of
the calibration data, is the threshold for turning data averaging
(smoothing) on or off. If the current exponentially-calculated oxygen
value is within the threshold value of the previous exponentially-
calculated oxygen value, the average of the two values is stored and
displayed. To set this parameter, proceed as follows:
You may now do one of the following:
• Press [N] as many times as necessary to select another Gas Value
Exp Filter menu option. Then, press
[Y] and proceed to the
appropriate section of this chapter for instructions.
• Press [N] until you reach the Done? option of the Gas Value Exp
Filter menu, then press
[Y] to return to the Calibrate System menu.
Gas Value Exp. Filter:
You have now completed the
programming of this submenu.
Set Window Threshold?
Gas Value Exp Filter:
Press [N] until this prompt
appears, then press
[Y] or [Enter].
Set Smooth Threshold?
Smooth Threshold Value:
To accept the current value, press
[Y] or [Enter]. To change the
current value, use the numeric
keys to enter a new value and
press
[Y] or [Enter] twice.
% [x.x]
Gas Value Exp Filter:
You have now completed the
programming of this submenu.
Set Smooth Threshold?