EasyManua.ls Logo

Getidy JN-23 - Sterilization Program Details

Default Icon
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Program sterilization sterilization drying vacuum
Name temperature time time times
134 04min 09min 1 times
134 04min 15min 3 times
134 18min 15min 3 times
121 30min 15min 3 times
121 20min 09min 1 times
134 or 121 03~60min 01~60min 1or 3 times
1.Solid
2.Hollow
3.Prion
4.Cotton
5.Plastic
6.User-
Definned
5.JN-23 steam sterilizer program introduction
Clickprogram(Fig.5-1) into the sterilization
program interface(Fig. 5-2) .
5.1 Sterilization program list
Fig.5-1 Fig.5-2
Fig.5-3
For example:PressHollow programicon,into the
interface(Fig.5-3).it is the interface befor the sterilization
program is running.
Running time.00:00:00is the time
each Stage.ALL:00:00:00is
the whole running.
Temperature.T1&T2is temperature inside
chamber.T3is temperature of heating
coil.T4is temperature of steam generator.
Vacuum times of this program.
Back to the previous interface
Sterilization temperature of this program.
Back to the main interface.
Name of sterilization program.
Water conductivity.
Pressure inside the chamber
Sterilization pressure of this program
Sterilization time of this program
Unlock icon
Start button.