Fig.5.3-13 Fig.5.3-14 Fig.5.3-15 Fig.5.3-16
5.3.1.10.AFTER STERILIZATION EXHAUST STAGE(Fig.5.3-13)
When the “sterilization time”countdown is over,the control system will open the exhaust-solenoid valve(Ev1) for exhausting untill pressure
downs to P=0.000bar.LCD shows “00:01:29”time countdown.
5.3.1.11.DRYING STAGE(Fig.5.3-14)
When the pressure inside the chamber downs to 0.000bar,the control system will open the vacuum pump-solenoid valve(Ev4) and start evacuating
and drying.LCD shows “00:14:59”time countdown.
5.3.1.12.BALANCE ATMOSPHERIC STAGE(Fig.5.3-15)
When the vacuuming and drying is over(the “00:14:59”time countdown is over).the control system will open the air-solenoid valve(Ev3) to balance
the pressure between in chamber and atmospheric.LCD shows “00:00:29”time countdown.
5.3.1.13.STERILIZATION RECORD PRINT STAGE(Fig.5.3-16)
After the balance atmospheric is over,the vacuum pump will be stoped.the control system will transfer sterilization report to external mini-printer and
U disk.
Fig.5.3-17 Fig.5.3-18
5.3.1.14.STERILIZATION EVALUATION STAGE(Fig.5.3-17 OR Fig.5.3-18)
When the sterilization working is over,the control system will evaluates the sterilization results(“PASS”or “FAILURE”).
WARNNING:
If the result of sterilization is “FAILURE”, medical
instruments of this time can not be used. And it's need to contact
professional engineer to check the machine.
Page 32 of 58