Fig.5.3-31 Fig.5.3-32
5.3.3 USER-DEFINNED PROGRAM
Fig.5.3-33 Fig.5.3-34 Fig.5.3-35
Click “User-definned”(Fig.5.3-33) enter into interface Fig.5.3-34.Press “ ”to change the number of “Temperature/
Vacuum times/Sterilization time/Drying time”.
Temperature range:121℃ or 134℃. Vacuum times range: 1 or 3. Sterilization time range:4~60. Drying time:1~60.
Then click“ ”to save and exit,enter into Fig.5.3-35 interface.Then press “Start”buttom to start.Refer to “chapter
”
5.3.1
or 5.3.2 PRE-VACUUM PROGRAM to check the running state diagram.
WARNNING:
If the result of sterilization is “FAILURE”, medical
instruments of this time can not be used. And it's need to contact
professional engineer to check the machine.
Page 35 of 58