E 00 Abnormally Exit
If need abnormally exit during sterilization,you have to press button,then click
ok ,LCD appear E00 alarm code and flashing.Press button again to relieve
Alarm,and start vacuuming-drying process.After 3 minute evacuating,the sterilizer will stop
working.This cycle is over,the operation display shows main interface.The door of sterilizer can
be opened.
E 98
Power outage
during operation
If you encounters abrupt power failure during the sterilizing operation,when the power is
restored, will appear on the LCD. Press the the alarm code button to clear this
Code.
Alarm Code Alarm Cause Solution
Teperature sensor
(steam generator)
is broken!
Measure:temperature sensor resistance value( ) 1000+(Ambient temperature *3.8)Ω
(1)if measurement result is OK,check poor contact point between the sensor and the
control board.(2)if measurement result is No,replacing the temperature sensor.
E 01
Teperature sensor
(heating coil) is
broken!
Measure:temperature sensor resistance value( ) 1000+(Ambient temperature *3.8)Ω
(1)if measurement result is OK,check poor contact point between the sensor and the
control board.(2)if measurement result is No,replacing the temperature sensor.
E 02
Teperature sensor
(inside chamber)
is broken!
Measure:temperature sensor resistance value( ) 1000+(Ambient temperature *3.8)Ω
(1)if measurement result is OK,check poor contact point between the sensor and the
control board.(2)if measurement result is No,replacing the temperature sensor.
E 03
APPENDIX 9:Alarm code and solutions of JN-23 Steam Sterilizer
Alarm Code
Alarm Cause
Release Alarm
Page 55 of 58