Mains frequency
Line 19 Authorization level: CD
Adaptation to the frequency of the electricity supply network.
1 = 50 Hz
2 = 60 Hz
Pump starting time
Line 20 Authorization level: CD
Setting always 0.0
Temperature indication (red LED)
Line 21 Authorization level: CD
Choose between two functions for the red LED on the panel:
01 = The LED is lit if the disinfection temperature has not been reached during the
program.
02 = The LED is lit during heating and the disinfection temperature has not been
reached.
Temperature, condensate cooling
Line 22 Authorization level: CD
Temperature above which condensate cooling must be activated.
0 - 99 in steps of 1 °C. Default value: 60 °C.
32 - 210 °F in steps of 1 °F. Default value: 140 °F.
Time to crack open
Line 23 Authorization level: CD
Time of activation of “door release” so that the door cracks open for 0-9.9 seconds.
Waiting time cracked open
Line 24 Authorization level: CD
Time for which the door must stay cracked open with a cracked-open time of 0-1800
seconds.
Run flushing program
Line 28 Authorization level: CD
Species which ushing program is to be called if the value of the parameter “ushing
program” in the ushing program instruction “Run ushing program” is 20.
0-20
Set flushing program
Line 29 Authority level: CDE
Sets which ushing programs are available on the respective buttons.
1-3