Function check
This must only be done by authorized personnel.
The machine is connected to the electricity supply and some components are
live.
Instructions, cable, switch
• Check that a goods placing sign and quick help instructions have been put up on
the wall behind the disinfector.
• Check that the isolator switch on the wall is working and that the connecting cable
is undamaged and free from defects.
Filters and valves
All supply lines must be closed when working on the pipe system.
• Check that check valves and manual shutoff valves are working properly.
• Check the lters and the ow limiters in the supply line (see under Cleaning the
lters). Clean if necessary.
• Check all pipe couplings. Tighten and seal if necessary.
Controls
Check the panel lights by:
• switching off the power with the isolator switch.
• and switching the power on again.
All the lamps on the panel must light up and the buzzer must sound.