Dosage, line 33
Line 33, sub-lines 01 - 04 Authorization level: BCD
Here you can set the dosing time for the various agents used during a process. The
various sub-lines correspond to the following agents:
Sub-line 01
Dosing time, descaler.
0 - 99 seconds in one-second steps.
Sub-line 02
Dosing time, rinse-aid.
0 - 99 seconds in one-second steps.
Sub-line 03
Dosing time for process agent.
0 - 99 seconds in one-second steps.
Sub-line 04
Dosing time, spray agent.
0 – 9.9 seconds in 0.1-second steps.
Empty container alarm, line 34
Line 34, sub-lines 01 - 05 Authorization level: CD
Here you can specify how the empty container alarms from the various containers are
managed.
The value consists of two or four digits:
00 = No alarm
0199 = 01 = an alarm must be given. The last two digits, 00 - 99, specify the number of
processes than can still be run after the warning.
02 = A fault alarm must be given. The machine cannot run until the alarm has been
acted on.
The various sub-lines correspond to the following agents:
01 = Alarm for descaler.
01.01 = Alarm for descaler processes remaining.
02 = Alarm for chemical agent.
02.01 = Alarm for chemical agent processes remaining.
03 = Alarm for rinse-aid.
03.01 = Alarm for rinse-aid processes remaining.
04 = Alarm for process agent.
04.01 = Alarm for process agent remaining
05 = Alarm for spray agent.
05.01 = Alarm for spray agent processes remaining.
Disinfection, lines 35 - 49
Normal disinfection method
Line 35 Authorization level: CD
There is a choice of four disinfection methods:
01 = D-T-D, Decentral-Thermal-Disinfection
02 = C-T-D, Central-Thermal-Disinfection
03 = D-C-D, Decentral-Chemical-Disinfection)