Standby disinfection method
Line 36 Authorization level: CD
There is a choice of four disinfection methods:
01 = D-T-D, Decentral-Thermal-Disinfection
02 = C-T-D, Central-Thermal-Disinfection
03 = D-C-D, Decentral-Chemical-Disinfection)
Lower disinfection temperature
Line 38 Authorization level: CD
The heating is switched on at this temperature.
Upper disinfection temperature
Line 39 Authorization level: CD
The heating switches off at this temperature. The upper disinfection temperature must
be 2 degrees above the lower disinfection temperature and above the disinfection
temperature.
Disinfection temperature
Line 40 Authorization level: CD
When this temperature is reached, the disinfection time in line 41 starts to count down.
Disinfection time
Line 41 Authorization level: CD
Time for thermal disinfection.
0 - 1800 seconds in one-second steps.
Max preheating time
Line 42 Authorization level: CD
Maximum time for preheating of the steam generator.
0 - 99 seconds in one-second steps.
Power sharing, steam generator
Line 43 Authorization level: CD
The steam generator has three elements. When this function is active, only two
elements are switched on when the water pump is on. This only applies to single-phase
machines.
00 = Not active
01 = Active
Cooling method
Line 44 Authority level: CDE
Choose a method for cooling the washed items after disinfection.
00 = Cooling method
01 = Global external cooling time
02 = Global extended external cooling time (after process)