Page 13 av 64
<Doc_TEC><Doc_503712400><Rel_B><Lang_EN>
Table of line numbers and line information
Line Sub-
line
Parameter Basic
setting
Setting range
Authorization
Lines 00-09: Machine-independent variables
0 - Authorization code 0000-9999 ABCD
1 - Checking and retrieving settings 1=Basic, 2=Factory, 3=Work CD
2 - Machine type FD1800, etc. -
3 01 Program version (main version) 00.=00-99 -
3 02 Program version (sub-version) .00=00-99 -
3 03 Program version (test version) 00-99 -
3 04 Program protocol 109 - -
4 - Serial number 0-999999 -
5 - Language 1 1=DE, 2=SE, 3=EN, 4=FR, 5=IT,
7=DK, 8=HU, 9=ES, 10=CZ,
11=PL
CD
6 01 Date, year 97 00-99 BCD
6 02 Date, month 01 01-12 BCD
6 03 Date, day 01 1-31 BCD
7 01 Time, hours 00 00-24 BCD
7 02 Time, minutes 00 00-59 BCD
8 - Demo 0 0 = Not active, 1 = Active CD
9 01 Address, interface 1 0=None, 1=RS-232, 2=RS-485 CD
9 02 Address, network address 1 0-9999 CD
Lines 10-29: Configuration
10 - Start-up code 0 0 = Not active, 1 = Active CD
11 - Child safety 0 0 = Not active, 1 = Active BCD
12 - Disinfection configuration 1 1=Normal, 2=Spare BCD
13 01 Dosing (button 2-3) 1 1=Manual, 2=Automatic BCD
14 01 Sound on door closing 1 0=None, 1=50 ms, 100 ms CD
14 02 Door type 1 1=Manual, 2=Automatic CD
14 03 IR sensor interval 1.0 1-9.9 sec DE
15 01 Acoustic alarm 0 0 = Not active, 1 = Active CD
16 01 Feedback activation 0 0 = Not active, 1 = Active DE
16 02 Feedback interval 0 0-255 DE
17 - Temperature scale 1 1=Celsius, 2=Fahrenheit CD
19 - Mains frequency 1 1=50 Hz, 2=60 Hz CD
20 - Pump starting time 0.0 0-5.0 seconds CD
21 - Temperature indication (Red LED) 1 1=Temperature error, 2=Heating CD
22 - Temperature, condensate cooling 60 0-99 °C CD
23 - Time to crack open 2.0 0-9.9 seconds CD
24 - Waiting time cracked open 0 0-1800 seconds CD
28 - Run flushing program 0 0-20 CD
29 - Set flushing program 1 1-3 CDE