<Doc_TEC><Doc_503712400><Rel_B><Lang_EN>
Page 14 av 64
Line Sub-
line
Parameter Basic
setting
Setting range
Authorization
Line 30: Program selection
30 11 Program selection button 1 1
00=No program, 01-20 Program no.
CD
30 12 Program name button 1 Economy program CD
30 21 Program selection button 2 2
00=No program, 01-20 Program no.
CD
30 22 Program name button 2 Normal program CD
30 31 Program selection buttons 2+5 4
00=No program, 01-20 Program no.
CD
30 32 Program name button 2 Normal+Detergent CD
30 41 Program selection button 3 3
00=No program, 01-20 Program no.
CD
30 42 Program name button 3 Intensive program CD
30 51 Program selection buttons 3+5 5
00=No program, 01-20 Program no.
CD
30 52 Program name button 3+5 Intensive program+Detergent CD
30 61 Program selection button 4 8
00=No program, 01-20 Program no.
CD
30 62 Program name button 4 Rim flushing CD
30 71 Set pedal 2 0-25 CDE
30 72 Name pedal 2 0-10 CDE
30 81 Set pedal + button 5 0 0-25 CDE
30 82 Name pedal + button 5 3 0-10 CDE
30 91 Program selection IR 2 CD
30 92 Program name IR 2 CD
30 101 Program name IR + button 5 0 CD
30 102 Program IR + button 5 3 CD
30 150 IR function 0 0=No IR, 1=Open/close door,
3=Start program
CD
Line 31: Interval disinfection
31 01 Interval disinfection 0 0 = Not active, 1 = Active CD
31 02 Interval disinfection programs 1 00=No program 01-20=Program
no.
CD
31 03 Limit for interval disinfection 9999 0-9999 program CD
31 04 Processes remaining to interval
disinfection
- 0-9999 program CD
31 05 Reset interval disinfection - 0=Zero BCD
Line no. 32: Inspection request
32 01 Processing of inspection requests 1 0=None, 1=Warning, 2=Stop CD
32 02 Acknowledgement of inspection
requests
- 0=Acknowledged CD
32 03 Counter type for inspection requests 1 1=Connection hours,
2=Number of programs
CD
32 04 Limit for inspection requests 9999 0-9999 CD
32 05 Counter for inspection request - 0-9999 -
Line 33: Dose size
33 01 Dosing time, descaling 6 0-99 seconds BCD
33 02 Dosing time rinse-aid 4 0-99 seconds BCD
33 03 Dosing time, process agent 30 0-99 seconds BCD
33 04 Dosing time spray agent 0.7 0-99 seconds BCD