Page 15 av 64
<Doc_TEC><Doc_503712400><Rel_B><Lang_EN>
Line Sub-
line
Parameter Basic
setting
Setting range Authorization
Line 34: Empty container alarm
34 01 Alarm descaling agent 0125 00=no alarm, 01.99=warning,
no. of programs after low level,
02=stop
CD
34 01.01 Alarm descaling agent
processes remaining
25 0-99 CDE
34 02 Alarm chemical agent 0200 00=no alarm, 01.99=warning,
no. of programs after low level,
02=stop
CD
34 02.01 Alarm chemical agent processes
remaining
0 0-99 DE
34 03 Alarm rinse-aid 0125 00=no alarm, 01.99=warning,
no. of programs after low level,
02=stop
CD
34 03.01 Alarm rinse-aid processes remaining 25 0-99 CDE
34 04 Alarm, process agent 0110 00=no alarm, 01.99=warning,
no. of programs after low level,
02=stop
CD
34 04.01 Alarm process agent processes
remaining
10 0-99 CDE
34 05 Alarm spray agent 0110 00=no alarm, 01.99=warning,
no. of programs after low level,
02=stop
CD
34 05.01 Alarm spray agent processes remaining 10 0-99 CDE
Lines 35-49: Disinfection
35 - Normal disinfection method 1 1=D-T, 2=C-T, 3=D-C-D,
4=C-K-D
CD
36 - Reserve disinfection method 1 1=D-T, 2=C-T, 3=D-C-D,
4=C-K-D
CD
38 - Lower disinfection temperature 91 80-95 ˚C CD
39 - Upper disinfection temperature 93 80-97 ˚C CD
40 - Disinfection temperature 85 80-95˚C CD
41 - Disinfection time 60 0-1800 seconds CD
42 - Max preheat time, steam generator 50 0-99 seconds CD
43 - Power sharing, steam generator 1 0=Not active, 1=Active CD
44 1 Cooling method 1 0=None, 1=Internal, 2=External,
3=Extended
CDE
44 2 Global external cooling time 120 1-180 sec CDE
44 3 Global extended cooling (after process) 120 0-180 sec CDE
45 1 Handling auto-disinfection 0 0=None, 1=Warning, 2=Warning
+ autostart
CDE
45 2 Auto-disinfection starts from 8 Time 0-23 CDE
45 3 Auto-disinfection permitted until 20 Time 0-23
46 - Disinfection to A0 0 0, 60, 600 CDE
48 - Dosage, chemical agent 70 0-99 seconds CD
49 - Chemical action time 0 0-600 seconds CD