Draining the machine
If the machine will not be used for more than 72 hours, Getinge Disinfection recom-
mends draining the steam generator and circulation pump.
Draining the steam generator
Unscrew the cover from the underside of the ball valve on the steam generator. Open the
ball valve and drain the steam generator. When it is empty, close the ball valve and put it
back and close the cover.
Draining the circulation pump
Unscrew the plug on the lower front edge of the pump and drain all the water from the
pump. Replace the plug and make sure it isn’t leaking.
Warm water can be found inside the steam generator and the pump
Ball valve
Plug
Circulation pump
Lid