<Doc_TEC><Doc_503712400><Rel_B><Lang_EN>
Page 28 av 64
Interval disinfection, line 31
Interval disinfection active
Line 31, sub-line 01 Authorization level: CD
Here you can activate or deactivate the interval disinfection function.
00 = Not active
01 = Active
Interval disinfection programs
Line 31, sub-line 02 Authorization level: CD
Choose the program to be run on interval disinfection. The programs, which are
described in detail under “Program selection”, have the following numerical values:
00 = No program
01 = Economy program
02 = Normal program
03 = Intensive program
04 = Normal program + dosing
05 = Intensive program + dosing
Limit for interval disinfection
Line 31, sub-line 03 Authorization level: CD
Set the number of programs performed between interval disinfections.
0 - 9999 in steps of one.
Processes remaining to interval disinfection
Line 31, sub-line 04
Number of programs to the next interval disinfection. The value cannot be changed but it
can be reset; see line 31, sub-line 05.
Reset interval disinfection
Line 31, sub-line 05 Authorization level: BCD
Zeroing of the counter of number of processes remaining to interval disinfection; see
line 31, sub-line 04.
See line 31, sub-line 04
00 = reset counter