EasyManuals Logo

Getinge FD1800 User Manual

Getinge FD1800
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
<Doc_TEC><Doc_503712400><Rel_B><Lang_EN>
Page 28 av 64
Interval disinfection, line 31
Interval disinfection active
Line 31, sub-line 01 Authorization level: CD
Here you can activate or deactivate the interval disinfection function.
00 = Not active
01 = Active
Interval disinfection programs
Line 31, sub-line 02 Authorization level: CD
Choose the program to be run on interval disinfection. The programs, which are
described in detail under “Program selection”, have the following numerical values:
00 = No program
01 = Economy program
02 = Normal program
03 = Intensive program
04 = Normal program + dosing
05 = Intensive program + dosing
Limit for interval disinfection
Line 31, sub-line 03 Authorization level: CD
Set the number of programs performed between interval disinfections.
0 - 9999 in steps of one.
Processes remaining to interval disinfection
Line 31, sub-line 04
Number of programs to the next interval disinfection. The value cannot be changed but it
can be reset; see line 31, sub-line 05.
Reset interval disinfection
Line 31, sub-line 05 Authorization level: BCD
Zeroing of the counter of number of processes remaining to interval disinfection; see
line 31, sub-line 04.
See line 31, sub-line 04
00 = reset counter

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge FD1800 and is the answer not in the manual?

Getinge FD1800 Specifications

General IconGeneral
BrandGetinge
ModelFD1800
CategoryCleaning Equipment
LanguageEnglish

Related product manuals