Machine-independent variables, lines 00-09
Authorization code
Line 0 Authorization level: ABCD
The machine has four authorization levels:
• Level A for users; display only (code 0000)
• Level B for users; change (code 2000)
• Level C for service personnel, change
• Level D for service personnel, critical settings.
Note: You can always view all level A program lines without the authorization code for
other levels. The code only protects the programming from unauthorized changes.
Checking and retrieving settings
Line 1 Authorization level: CD
Used to retrieve entire sets of settings (all line numbers) and to check which set is being
used. The following sets are available:
1 = Retrieve default settings
2 = Retrieve factory settings (the settings that were programmed before delivery)
3 = Use working setting
If a variable is changed after delivery of the machine, the value is automatically changed
from 02 to 03.
Machine type
Line 2
Shows the machine type to which the program applies.
1000 = Low spec (SP1000)
2000 = High spec (Getinge 2000)
This value cannot be changed.
Program version
Line 3, sub-lines 01-04
Shows the program version of the chips on the computer board.
The various sub-lines show:
01 = main version
02 = sub-version
03 = test version
04 = protocol version
These values cannot be changed.
Consecutive number
Line 4
Shows the serial number of the machine.
This value cannot be changed.