Handling auto-disinfection
Line 45 Authority level: CDE
This selection determines how auto-disinfection is handled.
01 = Selection of how to handle auto-disinfection
02 = What is the earliest time that auto-disinfection should begin
03 = What is the latest time that auto-disinfection should begin
Dosage, chemical agent
Line 48 Authorization level: CD
The chemical agent is a substance that is dosed into the tank in the nal phase of the
program. Here you can specify the time for which the dosing will run.
0 - 99 seconds in one-second steps.
Chemical action time
Line 49 Authorization level: CD
Here you can specify the time for which the chemical agent will act without it being
possible to remove the goods from the machine. The door is kept locked for the set time.
0 - 600 seconds in one-second steps.
Tank and water, lines 50 - 55
Tank level at rest
Line 50 Authorization level: CD
Specify the level to which the tank must be relled when the program is complete.
01 = low level (empty)
02 = intermediate level
03 = high level
Water level, filling at rest
Line 51 Authorization level: CD
Specify the type of water to ll the tank with when the process is nished.
The value consists of two digits:
01 = Cold water
02 = Hot water
03 = 03 means that both hot and cold water must be used (for mix ratio see line 52).
Mix ratio, water
Line 52 Authorization level: CD
Species the proportion of hot water as a percentage, where 100 = hot water only.