Page 29 av 64
<Doc_TEC><Doc_503712400><Rel_B><Lang_EN>
Inspection request, line 32
This function is used to get a continuous maintenance service of the machine. When the
function is active, a message stating that an inspection is due appears on the display after
a certain number of hours or program runs.
Processing of inspection requests
Line 32, sub-line 01 Authorization level: CD
The function can be enabled or disabled in two ways:
00 = Function disabled
01 = A warning message appears on the display when an inspection is due. The machine
can be run.
02 = A warning message appears on the display when an inspection is due. The machine
cannot be run.
Acknowledgement of inspection requests
Line 32, Sub-line 02 Authorization level: CD
Used by service technicians to remove the warning message and to zero the inspection
request counter. After this, the fault message must be acknowledged.
00 = acknowledgement of inspection requests.
Counter type for inspection requests
Line 32, sub-line 03 Authorization level: CD
The intervals between inspections can be dened in two ways:
01 = counts the time for which the machine has been connected to the power supply,
regardless of whether or not the machine has been working.
02 = Counts the number of programs run.
Limit for inspection requests
Line 32, sub-line 04 Authorization level: CD
Set the number of hours/programs run between inspection requests.
0 - 9999 in steps of one.
Counter for inspection request
Line 32, sub-line 05
Number of hours/programs to the next inspection request. This value cannot be changed.
It is set to zero when an inspection request is acknowledged; see line number 32, sub-
line number 02.