Insulation
Ball valve
O-ring
Cover
Clamping ring
Check valve
Gasket
Check valve
J155
5.3.1 Overheat protection
This must only be done by authorized personnel.
The machine is connected to the electricity supply and some components are
live.
The steam generator has an overheat cutout which trips if the temperature in the steam
generator becomes too high. The overheat cutout is t in the elements of the steam
generator. If the element overheats, the power is cut to the element and fault code F7
appears because the disinfection temperature has not been achieved. When the fault code
has been reset, the machine can be restarted. The steam generator is reset automatically.
To make sure that the overheat cutout does not trip immediately, the steam generator
should be allowed to cool down before a new process is started.
NOTE:
Before restarting the machine, always find out why the overheat cutout tripped.
The fuse may have tripped because the temperature sensor is faulty, because of incor-
rect operation, because of foaming additive or because the water supply is shut off or
blocked.