Semi-micro balance Mod. E50S/3
Page 9 of 38
Calibration is proformed in the configuration menu. Depending on the balance model, this may be done
externally, internally or automatically. (see chapter 6.3.5 "Calibration functions" and see chapter 12.3
"Calibration")
With the aid of the “Intelligent Calibration Mode” the balance can itself determine the size of the calibration
weight, which enables an exact calibration with different size weights (in 10 g, 50 g, 100 g and 500 g steps,
depending on implementation).
5.8 Dual Range
With the Dual Range balances, weighing is always first carried out in the fine range, which is 10 times more
precise. When the fine range is exceeded the balance switches automatically into the coarse range.
5.9 Standardized balance
Ask to the GIBERTINI Commercial Department.
5.10 Switching on the balance
Press «ON/OFF» to switch on the balance.
The balance carries out a diagnostic test in order to check the most important functions. After completion
of the start-up process (approximately ten seconds) “Zero” appears in the display.
The balance is ready for operation and is in the Weighing mode.
5.11 Auto-Standby Mode
The balance is equipped with an Auto-Standby mode, which can be activated or deactivated in the
configuration menu.
If the Auto-Standby mode is activated, the balance automatically switches to Standby some time after the
last weighing or key operation (current-saving function).
The delay before switching to Standby is defined in the configuration menu. (see chapter 6.3.6 "Weighing
mode")
Press any button or put on a weight in order to switch the balance from the Standby mode back to the
Weighing mode again.