8 Statistics
MODE OFF
STATISTIC
RECORDER
STAT./RECORDER
COUNT
100
Number of values to be stored automatically (1..500).
RECORDING
MANUAL
With «STO i» function key.
on a time basis.
after every weight change.
TIMEBASE
LOADCHANGE
TIMBASE
2.0
Time base for “storage” in seconds.
Functions of the statistics program and storage functions:
“MODE”
In this function-field you may define whether only the statistics-program, only the storage-program or both
programs simultaneously should be used.
“COUNT”
A number of measurements is laid down, after which automatic storage is to be terminated.
“RECORDING”
“MANUAL”: The «STO i» function key must be pressed for each value to be stored.
“LOADCHANGE”: The balance stores the measured value automatically after a load change.
“TIMEBASE”: The balance stores every value measured after a defined period (default: 2.0
seconds).
“TIMEBASE”
Definition of the time span for the recording of data in accordance with “RECORDING TIMEBASE” (for
example, every 2.0 seconds).
NOTE
In storing the first value a range of ±50% is determined. Subsequent values must be within this range
otherwise an error message will be issued.
Assignment of the function keys:
«
»
Take value, start/stop of automatic recording
«RES»
Before a new series of measurements the storage must be reset using «RES». The key
must be held down until the acoustic signal sounds and the measurement counter is
reset.
«
»
Permanently store data (only with recording activated).
«INF»
Change the display to the info display.
Displayed information:
- “Average value (AVERAGE)"
- “Standard deviation (STD DEV.)"
- “Relative standard deviation (STD DEV.-%)"
- “Maximum (MAX)"
- “Minimum (MIN)"
- “individual values"
Press «
» to exit the info display.