Semi-micro balance Mod. E50S/3
Page 33 of 38
12.2.1 Notes on correcting faults
The following table shows faults and their possible causes. If you cannot clear the fault using the table,
please contact a GIBERTINI Service Engineer.
Weight display does
not light
Balance not switched on
Connection to power adaptor is interrupted
Power supply has failed (interruption to current)
The power adaptor is defective
“OL” is shown in
display
The weight range has been exceeded (Observe information on the maximum
weight range)
“UL” is shown in
display
The weight range is below the range of the balance (Scale pan or scale pan
holder missing)
The weight display
fluctuates
continuously
The draft is too strong at the balance location
The balance support is vibrating or varying
The scale pan is touching a foreign body
The time chosen for “FLOATINGDISPLAY” is too short
The material being weighed is absorbing moisture
The material being weighed is being blown away or is evaporating
Strong temperature variations in the material being weighed
Results of weighing
are clearly incorrect
The balance was not correctly tared
The balance is not correctly levelled
The calibration is no longer correct
There are strong temperature variations
There is no display or
only dashes
The stability control (Balance functions) is set too sensitively
The time selected for “FLOATINGDISPLAY” is unsatisfactory
Configuration menu
cannot be changed
The password lock is activated in the configuration menu
The display flashes
continuously during
calibration
The balance location is not stable enough (Interrupt calibration with «ON/OFF»
and relocate the balance in a better position)
Use of an imprecise calibration weight (only applies to external calibration)
12.3 Calibration
The calibration of the balance is fixed in the Configuration menu. (see chapter 5.7 "Calibration of the
balance" and see chapter 6.3.5 "Calibration functions")
Possible types of calibration, depending on the model of balance:
External calibration by means of ICM (Intelligent Calibration Mode)
External calibration with freely selectable weight
Internal calibration
Automatic calibration
NOTE
The calibration can be interrupted at any time by pressing «ON/OFF».