Semi
-
Page 26 of 38
Display
Key
Step
+ 3,4770 g
STATISTIC
«»
Press until “STATISTIC” appears.
+ 3,4770 g
STO 0 RES
INF
«»
Record a stable measurement.
+ 3,4785 g
STO 1 RES INF
«»
Record a second measurement.
+ 3,4775 g
STO 2 RES INF
«»
Record a third measurement.
Retrieve the parameters:
Display
Key
Step
+ 0,0000 g
MITTEL 3,4777 g
«»
Show info display.
Average measurement.
+ 0,0000 g
STOR. 0,00076 g
«»
Standard deviation.
+ 0,0000 g
STOR.
-
% 0,02 %
«»
Relative standard deviation.
+ 0,0000 g
MAX 3,4785 g
«»
Maximum.
+ 0,0000 g
MIN 3,4785 g
«»
Minimum.
+ 0,0000 g
1 = 3,4770 g
«»
Measurement 1.
+ 0,0000 g
2 = 3,4785 g
«»
Measurement 2, etc.
+ 3,4775 g
STO 2 RES INF
«esc»
Exit info display.
Clear measurements:
Display
Key
Step
+ 3,4775 g
STO 2 RES INF
«»
Hold down the key until the acoustic signal sounds and
the measurement counter is reset.
+ 3,4775 g
STO 0 RES INF
Measurements cleared, balance ready for new statistics.
A statistics log is printed by pressing the «PRINT» key.