Tubing - Selection and Fitting
Introduction
There is a small Iso-Versinic® drain
tube inside the MINIPULS 3, with an
outlet at the base of the pump. The
tube, which limits the risks resulting
from a rupture of the peristaltic
tubing, is connected to the motor
assembly to prevent liquid from
seeping inside the instrument.
All other tubing is fitted to the
pump head. Gilson supplies two
types of tubing:
peristaltic tubing with two retaining stops,
connection tubing without retaining stops.
In normal operations, short lengths of peristaltic tubing are placed in
the pump head. With MP pump heads, it is essential that peristaltic
tubing with two retaining stops is used. The calibrated stops ensure
that the tubing is correctly tensioned when fitted to the head.
HF models can use either type of tubing. The user then links the
MINIPULS 3 with other hydraulic components of the system (for
example chromatographic columns) using connection tubing,
connectors, and couplers of appropriate diameter and material.