Troubleshooting
Problem Solution
Blank Display
If the display is blank when the pump is turned ON, proceed as
follows:
If the fan does not run.
Check the power connection and (if necessary) replace the power
supply. If the display remains blank, contact your local Gilson
representative.
If the fan runs.
Check the electronics by pressing the or key. If the pump
head does not rotate, there could be a problem with the electronics
(faulty board). Contact the service agency.
Dead Keypad
If the “+” and the “-” keys have some eect, the pump is probably
under remote control. If not, the keypad could need replacing.
Contact the service agency.
No GSIOC Response
Check the GSIOC ID number, GSIOC cables, connectors, and GSIOC
interface module. If these actions fail to correct the problem, there
could be a problem with the electronics (faulty board). Contact your
local Gilson representative.
Worn or Faulty Pump Head Replace the pump head.