To set the pre-determined values that are effective when the pump is
switched on, proceed as follows.
1 Put the pump in the “Autostart” mode (see Mode Selection).
2 Set the speed and direction, wait for five seconds, and then
switch off.
3 Switch on while pressing the STOP key, then select the “Hold
mode using the RABBIT key.
4 Press the STOP key to exit the configuration program.
The values set in this mode are preserved until the mode is changed
to either Autostart or Non-autostart.
Mode Selection
To select the mode:
1 Turn the power off.
2 Press the STOP key, and while holding it, turn the power on
again.
3 The display then shows one of three symbols to indicate the
current mode (“A”, “H”, or “-”) followed by a 2-digit identification
number (see Changing the Unit ID).
4 Press the RABBIT key until the symbol indicating the required
mode appears.
5 Press the STOP key to validate the selection.
47