user could set information about alarm sound and so on.
n ALARM VOL: selective from 1~7, 1 is the minimum volume, 7 is the maximum volume.
n ALM REC TIME: three options: 8 s, 16 s, 32 s.
n ALM PAUSE TIME: two options: 1 min and 2 min.
n ALM TYPE: UNLATCH.. "UNLATCH" refers to the situation that once the causes of alarm are eliminated, the
alarm will disappear automatically.
n KEYVOL: selective from 1~7 and OFF.
n ALM SOUND: it can be set as "OFF", and the symbol " " will appear on the screen.
The system will cancel the "OFF" of alarm sound in the following situations:
u The monitor is restarted;
u The alarm status is changed, for example, the system enters alarm pause status, or the alarm sound is
forbidden.
WA RN IN G
l Before starting monitoring, please check whether the settings of alarm limit are suitable for patient.
l Invalid alarm system may occur when setting the alarm limit to the limit value.
l When the alarm sound is turned off, the monitor will not make any sound even if a new alarm is triggered.
Therefore, user must carefully choose whether to turn off the alarm sound.
l In SILENCE or ALARM PAUSE status, set the alarm sound is as "OFF", then the system will
automatically terminate the status of SILENCE or ALARM PAUSE.
l When the alarm sound is "OFF", if the operator selects "SILENCE" or "ALARM PAUSE", the alarm
sound will be restored to the previous volume when it is turned off, and at this time, the system will enter the
status of silence or pause accordingly.
l Do not rely on the sound alarm system only for patient monitoring, user should pay close attention to the
patient's actual clinical situation.
NOTE:
l When alarm sound is turned off, a symbol " " will be displayed in technical alarm area.
l The alarm sound off is only valid when the device keeps turning on, once the device is restarted, this setup
to will be restored to the previous set value.
l The symbol " " means that the alarm sound is turned off, the system could not make any sound for the
alarm, so user must be careful when using this function. There are two ways to exit this status. Method 1:
Set the alarm sound as "ON" in the "ALARM SETUP". Method 2: Press the "SILENCE" button, the
symbol will become " ", then press the "SILENCE" button one more time, the system will return to