Chapter 14 TEMP Monitoring
14.1 Introduction
Two TEMP probes can be used together to obtain 2 temperature data, via comparing, the temperature difference can be
obtained.
14.2 Safety information
WA RN IN G
l Veri fy w he ther the probe cable is normal before monitoring. Unplug the temperature probe cable from the
socket, the screen will display the error message “T1/T2 TEMP OFF” and the audible alarm is activated.
l Take and place the temperature and cable carefully, and they should be rolled to loose loop when not used.
If internal electric wires are pulled too tight, the mechanical damage will appear.
l The calibration of the temperature measurement is necessary for every two years (or as frequently as
dictated by your Hospital Procedures Policy). When you need calibrate the temperature measurement,
contact the manufacture please.
14.3 Measurement
Measurement steps:
1. Select an appropriate TEMP probe according to patient type and measurement requirement.
2. Insert the probe cable into the TEMP jack directly.
3. Attach the TEMP probe to the patient properly.
4. Confirm that the alarm settings are suitable for the patient.
NOTE:
l Disposable TEMP probe can only be used once for one patient.
l The self-test of the temperature measurement is performed automatically once per 30s during the
monitoring. The test procedure lasts about 1s, which does not affect the normal measurement of the
temperature monitoring.
14.4 TEMP SETUP
Move the cursor to the TEMP hot key, then press the button to enter to "TEMP SETUP" menu.
n ALM REC: Select "ON" to enable report printing upon TEMP alarm.
n TEMP unit: or
n TEMP alarm setting
After entering the initial password "70808", the following contents can be set:
u ALM: pick "ON" to enable prompt message and data record during the TEMP alarm; pick "OFF" to disable
the alarm function, and prompt the symbol beside TEMP area.
u ALM LEV: set the alarm level, three options: HIGH, MED or LOW.
u Alarm for T1, T2 and TD occurs when the measured temperature exceeds set alarm high limit or falls below