Appendix B EMC Test Level Declaration - Guidance and Manufacture’s Declaration
Table 1: Electromagnetic emission
Guidance and manufacture’s declaration –electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The customer or the user should
assure that it is used in such an environment.
Emission test
compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class A
The Patient Monitor is suitable for use in all
establishments, other than domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emission IEC 61000-3-3
Complies