16.5 Influence factors
The following factors may affect the measurement accuracy:
n Leakage or internal leakage of sampling gas.
n Mechanical shock.
n Other interference sources of interference.
16.6 Alarm Information and Prompts
When the alarm switches are set to "ON" in relevant menus, the physiological alarms caused by the parameter
exceeding the alarm limit may possibly trigger the recorder to automatically output the alarm parameter value and
corresponding waveforms.
Tables below describe the possible physiological alarms, technical alarms and prompt messages occurring during CO
2
measurement.
Physiological alarms:
EtCO
2
measuring value is above upper alarm limit.
EtCO
2
measuring value is below lower alarm limit.
InsCO
2
measuring value is above alarm limits.
AwRR measuring value is above upper alarm limit.
AwRR measuring value is below lower alarm limit.
RESP stops(in specific time interval, no RESP can be detected
using CO
2
module.).
Technical alarms:
Check that the sensor is properly plugged
in. Reinsert or reset the sensor if necessary.
If error persists, return sensor to factory for
servicing.
The sensor temperature is
greater than 40℃
Make sure the sensor is not exposed to
extreme heat. If error persists, return sensor
to factory for servicing.
The atmospheric pressure
exceeds the specified range.
Check whether the sampling cannula is
occluded or twined.
An error was found during
Zeroing
Check and clean the air adapter, if the error
is not corrected, please calibrate zero.
The value being calculated is
greater than the upper CO
2
limit.