Single PVC not belonging to the type of above mentioned
PVCs.
5 consecutive QRS complex , RR interval is less than
0.5s.
5 consecutive QRS complex, RR interval is longer than
1.5s.
When HR is less than 100 beats/min, no heart beat is
tested during the period 1.75 times of the average RR
interval; or When HR is larger than 100 beats/min, no beat
is tested with 1 second.
No QRS complex and pacing pulse are available during
the period 1.75 times of the average R-R interval (only
considering patients with pacemaker.)
When pacing pulse is available, no QRS complex exists
during the period 1.75 times of the average RR interval
(only considering patients with pacemaker.)
Applicable patient type: "All patients" refers to perform Arr.analysis on patients either with pacemakers or without
pacemakers.
"Without pacemaker": refers to perform Arr. Analysis only on the patients without pacemakers.
"With pacemaker": refers to perform Arr. Analysis only on the patients with pacemakers.
Prompt message:
The QRS template building required for Arr. Analysis is in process.
NOTE:
l Arrhythmia name displays in the alarm area.