78
different default configurations, as shown in the following table.
Default configurations
Default inflation value
(mmHg/kPa)
Selectable inflation value in manual mode in
NIBP menu(mmHg/kPa)
FACTORY DEFAULT ADU
CONFIG
150
80/100/120/140/150/160/180/200/220/240
FACTORY DEFAULT PED
CONFIG
100
80/100/120/140/150/160/180/200
FACTORY DEFAULT NEO
CONFIG
70
60/70/80/100/120
USER DEFAULT ADU CONFIG
150
80/100/120/140/150/160/180/200/220/240
USER DEFAULT PED CONFIG
100
80/100/120/140/150/160/180/200
USER DEFAULT NEO CONFIG
70
60/70/80/100/120
Press "MENU" button to enter "SYSTEM MENU" menu, then select a factory or user configuration in "DEFAULT"
menu, after configuration, return to the main interface to select NIBP hot key to enter "NIBP SETUP" menu. Here the
initial value for "Inflation" is the initial inflation pressure value corresponding to default configuration, as shown in the
above table. Move the cursor to the "Inflation" item and press it, inflation value range(as shown in the above table) in
MANUAL mode can be seen.
NOTE:
l "Inflation" is used to help user select the cuff inflation pressure next time, but the subsequent inflation is
the measurement value of last systolic pressure based on the same patient. The system momorizes the value,
which can shorten the measurement time of the same patient and increase the measurement accuracy.
l If user only sets the "Patient type" in "PATIENT SETUP" interface, does not perform any selection in
"DEFAULT", the system will operate according to the initial setting of relative module parameter in
"Patient type". The change of default type setting in "DEFAULT" will alter the "Patient type" in
"PATIENT SETUP" interface.
n NIBP alarm setting
Enter the initial password "70808" to set the following contents:
u AlM: when pressure alarm occurs, the system will prompt and store the alarm information after
selecting "ON", it will not alarm when selecting "OFF", and " " will appear in parameter area.
u ALM LEV: HIGH, MED and LOW, "HIGH" represents the most serious alarm.
u Pressure alarm is set according to the HIGH and LOW limits, alarm is activated when the pressure is higher
than the HIGH limit or lower than the LOW limit. Alarm for systolic pressure, mean pressure and diastolic
pressure can be set separately.
n RESET
Restore measurement status of the pressure pump. Press this button to restore the initial settings of the pressure pump.
When the pressure pump does not work properly and the system fails to give prompt information for the problem, press