Neonate: (low limit+1)~135
Low limit of systolic pressure
High limit of diastolic
pressure
Adult: (low limit+1)~270
Pediatric: (low limit+1)~150
Neonate: (low limit+1)~215
Low limit of diastolic
pressure
High limit of mean pressure
Adult: (low limit+1)~235
Pediatric: (low limit+1)~165
Neonate: (low limit+1)~110
Low limit of mean pressure
A.12 SpO
2
Note:The claims of SpO
2
accuracy shall be supported by clinical study measurements taken over the full
range. By artificially inducing to different oxygen levels, in the range of 70% to 100% SaO
2
, compare the
SpO
2
values collected by the secondary standard pulse oximeter equipment and the tested equipment at the
same time to form data pairs, which are used as an accuracy analysis. Because pulse oximeter equipment
measurements are statistically distributed, only about two-thirds of pulse oximeter equipment measurements
can be expected to fall within ±Arms of the value measured by a CO-OXIMETER.
Measurement and display
range
70~100 %: ±2 %;
0~69 %: unspecified;
±2 bpm or ±2%, whichever is greater