connection between the electrode and
lead cable.
Patient suffers from BRADY.
Check the patient' condition; Check the
connection between the electrode and
lead cable.
Patient suffers from Arr. of VT>2.
Check the patient' condition; Check the
connection between the electrode and
lead cable.
Patient suffers from Arr. of
MISSED BEATS.
Check the patient' condition; Check the
connection between the electrode and
lead cable.
The pacemaker is not paced.
Check the connection of the pacemaker;
check the connection between the
electrode and lead cable; check the
patient' condition.
No pacemaker signal is captured.
Check the patient' condition; Check the
connection between the electrode and
lead cable.
"ECG LEAD OFF or
RESP LEAD OFF"
RL lead of ECG is not connected
correctly or ECG lead is not
connected correctly.
Check the connection of RL lead or
ECG lead cable.
The V lead of ECG is not
connected correctly.
Check the connection of V lead.
The LL lead of ECG is not
connected correctly.
Check the connection of LL lead.
The LA lead of ECG is not
connected correctly.
Check the connection of LA lead.
The RA lead of ECG is not
connected correctly.
Check the connection of RA lead.
At least one lead among RA, RL
and LL is not connected correctly.
Check the connection of RA, RL or LL
lead.
SpO
2
sensor is not connected
correctly.
Check the connection of SpO
2
sensor.
The pulse signal of the patient is
too small so that the system cannot
perform pulse signal analysis.
Check the connection of sensor; Check
the patient' condition.