A-1
A
TYPICAL OPERATING
PARAMETERS
Typical operating parameters for the SHIPSURVEYOR
instrument are as follows:
Gas Range Resolution Accuracy
Range
LEL 0 to 9.9% 0.1% *
10 to 100% 1%
+
2% /
+
5% of reading
VOL 0 to 9.9% 0.1%
10 to 100% 1%
+
1% /
+
5% of reading
O
2
0 to 20.9% 0.1%
+
0.1% /
+
5% of reading
21 to 25% 1%
+
5% of reading
H
2
S 0 to 100ppm 1ppm
+
1ppm /
+
5% of reading
CO
2
0 to 5% 0.1%
+
0.1% /
+
5% of reading
5 to 20% 0.5%
+
5% of reading
CO 0 to 1000ppm 1ppm
+
5ppm /
+
5% of reading
* Congurable option, see page 2-2.
Notes:
All the values above are at normal temperature and pressure.
Humidity is between 0% and 100% RH (non-condensing).
Pressure changes at the inlet and exhaust should be minimised as they
may cause transient changes in reading.