OPERATOR MAINTENANCE
5.2 Filter Replacement
5.2.1 Internal Filter
The SHIPSURVEYORinstrumenthasanintegralhydrophobic
ltertted.Toreplacethislter,refertoMaintenanceManual
(GMIPartNo.48166).
5.2.2 Probe Handle (Accessory) - Part No. 48120
Hydrophobicandcottonparticulateltersintheprobehandle
minimise the danger of water and dust ingress.
Toreplacethelter(s),proceedasfollows:
1) Unscrewtheprobehandleassembly.
HYDROPHOBIC FILTER
PROBE HANDLE ASSEMBLY
COTTON PA RTICULATE FILTER
2) Removethecottonparticulatelteranddiscard.
3) Removethehydrophobiclter.
4) Clean the probe handle to make sure that it is free
from dirt and water.
5) Fitanewcottonparticulatelter.
6) Fit the hydrophobic lter.The yellow label on the
filter fits against the yellow label on the probe
handle.