OPERATOR MAINTENANCE
5.2 Filter Replacement
5.2.1 Internal Filter
The SHIPSURVEYORinstrumenthasanintegralhydrophobic
ltertted.Toreplacethislter,refertoMaintenanceManual
(GMIPartNo.48166).
5.2.2 Probe Handle (Accessory) - Part No. 48120
Hydrophobicandcottonparticulateltersintheprobehandle
minimise the danger of water and dust ingress.
Toreplacethelter(s),proceedasfollows:
1) Unscrewtheprobehandleassembly.
HYDROPHOBIC FILTER
PROBE HANDLE ASSEMBLY
COTTON PA RTICULATE FILTER
2) Removethecottonparticulatelteranddiscard.
3) Removethehydrophobiclter.
4) Clean the probe handle to make sure that it is free
from dirt and water.
5) Fitanewcottonparticulatelter.
6) Fit the hydrophobic lter.The yellow label on the
filter fits against the yellow label on the probe
handle.