EasyManua.ls Logo

Gossen MetraWatt ECUTEST ST BASE - Page 26

Gossen MetraWatt ECUTEST ST BASE
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
26 Gossen Metrawatt GmbH
(De)activate the QEDIT function (quick
edit – with SECUTEST DB COMFORT
only – Z853S or feature KD01).
If QEDIT is activated, you can fill in
additional fields for the test object in
the next step.
Confirm with the green checkmark.
8 If you activated the QEDIT function in the
previous step, you can now fill in all the
test object’s fields. Confirm with the
green checkmark.
9 An overview of the database
appears along with the newly
created test object.
Press the save icon (floppy disk) in
order to store the measurement results.
A message appears indicating that the
data have been successfully saved and
the display is switched to the measuring
view.
Note
The storage process can be aborted
by pressing the ESC key. The display
is returned to the memory menu. All
measured values can be deleted by
once again pressing the ESC key.
Note
You can send the test results to a PC
on which IZYTRONIQ software is run-
ning. This function is known as
“push-print” and can be implemented
via USB or Bluetooth
®
.
Database extension SECUTEST DB
COMFORT (Z853S or feature KD01)
and, if applicable, feature M01 (Blue-
tooth
®
) are required to this end.
Complete information regarding
push-print and a description of the
application can be found in
IZYTRONIQ online help.