Gossen Metrawatt GmbH 29
SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
13.5 RINS – Insulation Resistance
Measurement for Protection
Category II DUTs
U+ = increase UISO(set)
U– = decrease UISO(set)
➭ Connect the DUT to the test socket.
➭ Contact all conductive, exposed parts
with test probe P1.
Meas. Parameter Meaning
Measurement type
LN(TS) – P1 Testing is conducted between
short-circuited LN mains termi-
nals at the test socket and test
probe P1.
(connection via test socket, via
VL2E, AT3-IIIE,
AT3-IIS, AT3-II S32
or AT16DI/AT32DI adapter
)
UISO(set) U+/U–
> 50 ... < 500 V Variable test voltage can be en-
tered with the numeric keypad
Select measuring function
Save measured values to clipboard
Save measurements under ID no.