SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
40 Gossen Metrawatt GmbH
13.16 EXTRA – Special Functions
SECUTEST ST BASE(10)
If a QR code is displayed:
Scanning the code makes it possible to
download the current operating instructions
from www.gossenmetrawatt.com, which
can be read at a tablet.
FeatureI01 (e.g. SECUTEST ST PRO &
SECULIFE ST BASE(25) )
In this case, the addi-
tional functions are
assigned to the rotary
switch’s EXTRA position.
Select the desired measuring function.
Temp. – Temperature Measurement
Temperature measurement is conducted
with either a Pt100 or a Pt1000 temperature
sensor – the sensor type is automatically
detected internally.
Select measuring function
1
EXTRA
2
Select measuring function
1
EXTRA
Temp. or IZ
2
3
Connect the DUT
4
V
COM
Start test
5
Save measured values to clipboard
6
Stop test
7