SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
32 Gossen Metrawatt GmbH
13.8 IE – Device Leakage Current
1
Connection also via AT16DI/AT32DI (only sensible with differential
method)
2
Measurement must be performed with mains polarity in both direc-
tions. The largest value is documented.
Before conducting any leakage current measure-
ments, make sure that the “Ref. voltage L-PE” and
“Alt. Test Freq.” measurement parameters have been
set correctly in SETUP (see section 12.4).
Connect the DUT to the test socket.
Contact accessible, conductive parts
which are not connected to the protec-
tive conductor with test probe P1.
For DUTs with applied parts: Additionally
contact the short-circuited inputs for the
applied parts with test probe P1.
Measuring
Parameter
Meaning
Measurement type
Direct
Direct measuring method (via test
socket
1
), optional probe contact
Differential Differential current measurement
(via test socket)
Alternative Equivalent leakage current mea-
suring method with probe contact
(via test socket, AT16DI/AT32DI)
AT3 adapter
Only test instruments with feature
I01: Measurement with AT3-IIIE,
AT3-IIS or AT3-II S32 adapter
See section 13.19
Clamp
Only test instruments with feature
I01: See section 13.18
Polarity
2
– only with measurement type
direct, differential and AT3 adapter
Normal Selection of polarity for mains
voltage to the test socket
Reversed
Select measuring function
1
IE
Select parameters
2
Set parameters
3
Connect the DUT
4
Direct / Diff / Alt
P1
Start test
5
Acknowledge line voltage warning
6
!
Direct & differential & AT3 adapter:
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9