Gossen Metrawatt GmbH 33
SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
13.9 IA – Leakage Current from the Applied
Part
Before conducting any leakage current measure-
ments, make sure that the “Ref. voltage L-PE” and
“Alt. Test Freq.” measurement parameters have been
set correctly in SETUP (see section 12.4).
➭ Connect the DUT (to test socket).
➭ Contact the short-circuited inputs for the
applied parts with test probe P1.
Meas. Parameter Meaning
Measurement type
Direct P1 Direct measuring method
(via test socket) with test probe P1
Alternative P1 Equivalent leakage current mea-
suring method (via test socket)
with test probe P1
Perm. conn. P1 Permanently installed DUT
Phase angle – for direct (P1) & perm. conn. (P1)
only
0° or 180°
Selectable phasing for the internal
generator relative to mains phasing
Polarity – only with measurement type
direct, differential and AT3 adapter
Normal Selection of polarity for mains
voltage to the test socket
Reversed
Select measuring function
Direct P1 /
P1
PC 1 device under test
Acknowledge line voltage warning
Save measured values to clipboard
Save measurements under ID no.