13.12 U – Measuring Voltage
(only with feature I01, e.g. SECUTEST ST PRO and
SECULIFE ST BASE):
Attention!
Use only the included, contact-pro-
tected KS17-ONE measurement
cables when measuring dangerous voltage.
Special Case: Permanently Installed DUT
➭
When testing mains power packs or chargers:
Connect the DUT’s mains plug to the test
instrument’s test socket.
➭ Connect the DUT’s output, e.g. for mea-
suring safety extra-low voltage, to the V
and COM sockets.
Meas. Parameter Meaning
Measurement type
V – COM TRMS value + AC + DC for per-
manently connected DUTs
V – COM (with mains) TRMS value + AC + DC,
with mains to test socket
Polarity
Normal
Reversed
Selection of polarity for mains volt-
age to the test socket
(only with V – COM (with mains))
Select measuring function
V – COM (with mains)
Measuring voltage
VCOM
V – COM
Measuring
VCOM
Voltage
Acknowledge line voltage warning
!
Only for meas. type:
(with mains)
Save measured values to clipboard
Save measurements under ID no.