SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
14.2 Example of an Integrated Test
Sequence
Individual test steps can be configured with the sequence
parameters (see operating instructions).
* If the settings of the classification parameters are de-
tected automatically, they’re identified by an orange
border (in this case and ). However, they have
to be entered manually if they’re not automati-
cally detected, or if they’re detected incorrectly.
Connect the DUT to the test instrument
in accordance with the selected test se-
quence.
– Test socket
– Permanent connection
– Adapter
Connection depends on the type of DUT.
For testing extension cords in accordance
with standards: connection to the test
socket via the following adapter
EL1: for single-phase extension cords
VL2E/AT3-IIIE: for single and 3-phase
extension cords
The following checks are run auto-
matically before the test sequence is
started:
Probe check P1 (determines whether or
not test probe P1 is connected and fuse
link P1 is intact)
Insulation test (whether or not the DUT is
set up in a well-insulated fashion)
On test and short-circuit test.
In order to be able to detect a short-cir-
cuit at the DUT, testing is conducted
between L and N, as well as LN and PE.
If you’ve set the “
Detected classification
parameter for the respective test
sequence to “Always accept” and the
Auto-detection of
” parameter to “Con-
nection and SK” (before triggering
Start
), the
following additional checks will be run before
the test sequence is started:
Meas. Parameter Meaning
Standard Test standard / extension cord
Protection category * PC1/PC2/PC3
Connection type * Test socket / permanent /
adapter
Measurement type (MA)
*
Active or passive DUT
(on test:
on = passive, off = active)
Detected classification No auto-detection:
all classification parameters
such as connection, protection
category and measurement type
must be entered manually.
Always accept:
All classification parameters ac-
tivated under “Auto-detection of”
are detected automatically and
accepted.
Select test sequence
1
A3
Initial page
2
abc
Set sequence parameters
3
Set classification parameters
4
a
b
c
Auto-detection of Any desired combinations for
automatic detection of:
– Connection
– Protection category (PC)
– Measurement type (MT)
Meas. Parameter Meaning
b
a
c
a
b
Connect the DUT
5
Checking Connection and Starting the Test
6