SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
46 Gossen Metrawatt GmbH
Evaluation Procedure
During a measurement procedure, evalua-
tion is performed automatically for some test
steps within a test sequence, while for oth-
ers it must be performed manually:
Case A – automatic triggering of evaluation:
Evaluation (with a duration of, for exam-
ple, 5 seconds) is started automatically
as soon as the measured value has sta-
bilized. The worst value which occurs
during this duration is saved, and auto-
matic switching to the next test step
ensues.
Case B – manual triggering of evalua-
tion:
Evaluation is started after pressing the
measurement value recording icon (dis-
play: 0). After a specified period of time
has elapsed, the worst value is saved to
the right of wc: (worst case), and the
number 1 is displayed in the measure-
ment value recording icon indicating that
the first measured value has been saved.
Pressing the measured value recording
icon again restarts the evaluation proce-
dure. If the worst value is worse than the
value obtained for the previous measure-
ment, the new value is used. However, if
this value is better than the previous
worst value, the original value remains in
the display. Depending on whether you
want to delete the last value saved to the
clipboard or all values, press the
icon an appropriate number of times.
Proceed to the next test by pressing
the icon shown at
the right.
Attention!
!
Danger of electric arcs and damage
to surfaces (feature G01 or G02
only).
High test current is applied during
test steps of sequence parameter
RPE IP with 10 or 25 A test current.
It’s activated as soon as the evalua-
tion period starts and remains active
until the evaluation period has ended.
Maintain contact between the probe
and the DUT for the entire duration of
the evaluation!