Gossen Metrawatt GmbH 39
SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
13.15 EL1 – Function Test for Extension
Cords
This function permits an evaluation of
the continuity of the active conduc-
tors L(1, 2, 3) and N of an extension
cord. The PE conductor isn’t tested
in this case!
See corresponding single measurements for
the testing of R
PE
and R
INS
.
See section 14 with regard to testing
extension cords per DIN VDE 0701-
0702, for which R
PE
and R
INS
are
measured.
Connecting the EL1 Adapter
➭ Connect the EL1 adapter to the P1 spe-
cial sockets at the test instrument.
➭ Connect the plug at the end of the ex-
tension cord to the test socket.
➭ Connect the coupling socket at the end
of the extension cord to the plug at the
EL1 adapter.
Connecting Test Adapters VL2E and AT3-IIIE
➭
Connection examples can be found in section
13.19.
Continuity test for L and N
Meas. Parameter Test for
Measurement type
Continuity
L(1/2/3), N
Short-circuit
between
L(1/2/3), N
Reversed polarity /
clockwise
phase sequence
EL1 adapter
XX—
EL1 adapter
(continuity only)
X——
VL2E adapter
XXX
AT3-IIIE adapter
XXX
Select measuring function
Save measured values to clipboard
Save measurements under ID no.