SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
38 Gossen Metrawatt GmbH
13.14 P – Functions Test
The following connection types are possible:
Test socket
CEE adapter (only for connection via
single-phase CEE or “caravan socket”)
AT3 adapter (AT3-IIIE, AT3-IIS, AT3-IIS32)
AT16DI/AT32DI
Note
These or similar adapters can be
used for the function test (initial
startup of the DUT), but measure-
ment of apparent and active power,
power factor and current consump-
tion is only possible when the DUT is
directly connected to the test socket
or via the CEE adapter (single-phase
CEE socket only).
Connect the DUT to the test socket.
Meas. Parameter Meaning
Polarity
Normal Selection of polarity for mains
voltage to the test socket
Reversed
Select measuring function
1
P
Select parameters
2
Set parameters
3
Connect the DUT
4
Start test
5
Acknowledge line voltage warning
6
!
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9