4-4
Specifications
External Input/Output Functions
Control Software
Item Description
Input/output types • Trigger input (1 ch)
• Logic input (4 ch) or Pulse input (4 ch)
• Alarm output (4 ch)
*Switch between Logic and Pulse
Input specifications Terminal type : M4 screw type terminals
Maximum input voltage : 0 to 24 V (single-ended ground input)
Input threshold voltage : Approx. 2.5 V
Hysteresis : Approx. 0.5 V (+ 2.5 to + 3 V)
Alarm output specifications Output format : Relay contact output (NO/NC)
Terminal type : M4 screw type terminals
Rated : 250 VAC/2A
Output conditions : Level judgment, window judgment, logic pattern judgment,
pulse judgment, rate of change judgement
*Set output conditions of four channel values.
Pulse input Revolutions mode (engines, etc.)
Function : Counts the number of pulses per second; enables them to be converted to rpms
Span : 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M RPM/F.S.
Counts mode (electric meters, etc.)
Function: Displays a count of the number of pulses for each sampling interval from
the start of measurement
Span : 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M C/F.S.
Inst. Mode
Function: Counts the number of pulses for each sampling interval
Resets the count value after each sampling interval
Span : 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M C/F.S.
Maximum number of pulse input
Counts, Inst. Modes : 50 k/sampling interval
Revolution counts mode : 50 k/s
Item Description
Compatible operating system Windows 2000, XP, Vista (32-bit version, 64-bit version)
Functions Main unit control, realtime data capture, data conversion
Allowed connection Up to 10
Number of channels per connection
10 ch maximum
Maximum number of channels 100 ch maximum
Settings AMP settings, data settings, trigger/alarm settings, report settings, others
Captured data Realtime data (CSV, Binary)
Memory data (CSV, Binary)
USB memory data (CSV, Binary)
Display Analog waveforms, logic waveforms, pulse waveforms, digital values
Display modes Y-T View, X-Y View, Digital View, Meter View, Report View
File conversion Between cursors, All data
Monitor functions Alarm monitor enables sending of email to the specified address
Dual-screen function Displays the current data alongside past data
Statistic/History Displays maximum, minimum, and average values during measurement
Report function Enables creation of daily or monthly files