2.4 Installimg and Connecing MT100
Installation Site
Install MT100 in an indoor location that meets the following conditions.
Instrumentation panel
MT100 is designed for installation in an instrumentation panel.
Attach the desktop case (B-541) when using the device on a desk.
Locations that allow adequate ventilation
Install the MT100 in a place where adequate ventilation is provided to allow heat dissipation.
To install multiple devices, refer to page 4-9 for panel cut-out dimensions.
Make sure adequate space is provided when installing other devices close to the MT100 by following
the specified panel cut-out dimensions.
Locations not subject to mechanical vibration
Install the MT100 in a location not subject to excessive mechanical vibration.
Flat surface
Make sure the MT100 is installed horizontally.
(Mounting the device with a 0 to 30 degree rearward inclination angle is allowed.)
CAUTION
Condensation may occur when the device is moved from a low to high temperature/humidity
environment or exposed to a significant temperature change.
Using the MT100 with thermocouple input may cause measurement error.
In such cases, leave the device for at least one hour to warm it up to room temperature before you start
any operation.
Do not install the MT100 in the following locations.
Outdoors
Locations exposed to direct sunlight or near heating equipment
Choose an environment that is close to 23
°C
and avoid places with significant temperature changes.
Locations subject to excessive heavy fumes, steam, humidity, dust, or corrosive gas
This device may malfunction or be damaged by heavy fumes, steam, humidity, dust, or corrosive gas.
Do not install the MT100 in locations subject to such conditions.
Locations subject to electromagnetic interference
Do not place the MT100 near magnets or instruments that emit electromagnetic radiation.
Using this device close to a strong electromagnetic field may cause measurement error.
Installation Procedure
Use a steel plate (thickness must be within 2 to 26 mm) for the mounting panel.
Use the two brackets supplied with the device to mount the panel.
There are two sets of brackets. These brackets must be attached on two opposite faces: top and bottom OR
left and right surfaces.