EasyManua.ls Logo

GRAPHTEC MT100 - Page 31

GRAPHTEC MT100
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-14
Checks and Preparation
Logic/Pulse Specifications
*Switch between logic and pulse input.
Trigger Input Specifications
Alarm Output Specifications
Item Description
Number of output channels 4 channels
Output format Relay contact output (NO/NC)
<relayratings>
contact rating : 250 VAC/2A, 30 VDC/2A
contact relay life : 100,000 operations
Withstand voltage Between output and GND: 1000 VAC/1 minute
Item Description
Number of input channels 4
Input voltage range 0 to +24V max. (single-ended ground input)
Threshold level Approx. +2.5V
Hysteresis Approx. 0.5 V (+2.5 to +3 V)
Item Description
Number of input channels 1
Input voltage range 0 to +24V max. (single-ended ground input)
Threshold level Approx. +2.5V
Hysteresis Approx. 0.5 V (+2.5 to +3 V)

Table of Contents