3-23
Settings and Measurement
(2)-2 Backup Settings
MT100 provides a back up function where captured data can be backed up periodically (see figure below).
This section allows you to set data back up conditions.
(2)-3 Statistical Calculation
Two calculations can be performed in MT100.
This section allows you to set details of the statistical calculation.
CHECKPOINT
• Operation results are displayed in the Digital + Calculation Display screen.
•
Calculation will start upon power ON. Pressing the Start key to begin measurement will clear the calculation.
AutomaticbackuptoFTPserver
MT100 FTPserver
USBmemor
Automaticbackup
toUSBmemory
Measurementinput
Ethernet
Internalmemory
Setting items Description
Backup Intervals Sets the interval to perform back up of captured data.
Off, 1, 2, 6, 12, 24 h
Backup Desination Sets the location where back up data will be stored.†
USB1 Data is backedup in the USB memory.
This is enabled only when data is captured in the internal Flash memory.
FTP Data is backed up in the FTP server in the network.
* FTP settings in the FILE menu is required (see 3-32).
Folder Name Specifies the folder where data will be saved.
ex) \GRAPHTEC\TEST\20071204
Setting items Description
Calc. Settings 1 (2) off Calculation is not performed.
Average Displays the simple average value during data capture.
Max Displays the maximum value during data capture.
Min Displays the minimum value during data capture.
Peak Displays the peak value data during data capture.
RMS Displays the effective value of the data during data capture.
The following formula is used.
*
D : data n : number of data
R.M.S
ΣD
2
/