Modbus Protocol
Address 110 (DHCP Flag) (Format: 16 Bit Word)
Get the value that states the instrument's TCP/IP connection type i.e. Either Manual or
Auto DHCP. The DHCP Flag consists of 1 word which refers to 0 = Manual and 1 = Auto
DHCP.
Address 111 (Language Identification code) (Format: 16 Bit Word)
Get / Set the value that states the instrument's language type. The Language
identification code consists of 1 word which refers to 0 = English, (Future implementation:
1 = French, 2 = German, 3 = Italian, 4 = Spanish, 5 = Chinese Simplified and 6 =
Japanese.)
Address 112 (Mode of Operation) (Format: 16 Bit Word)
Get / Set the instrument's Mode of Operation. The Mode of Operation consists of 1 word
which refers to 0 = Offline, 1 = Online TOC, 2 = Online Conductivity, 3 = Digital TOC, 4 =
Manual TOC, 5 = Manual Bottle, 6 = Manual Clean, 7 = Manual Diagnostics and 8 =
Standby.
Note: Modes 5 = Manual Bottle and 7 = Manual Diagnostics can only be read and not set.
Address 113 (Current State) (Format: 16 Bit Word)
Get the value that refers to the instrument's current operational state. The Current State
consists of 1 word which refers to 0 = Idle, 1 = Flushing, 2 = Analyzing, 3 = Clean, 4 =
Standby, 5 = Offline, 6 = Error, 7 = Bottle mode and 8 = Diagnostic mode.
Address 114-115 (Cycle Time) (Format: Time 1)
Get / Set the instrument's Cycle Time. The Cycle Time consists of 2 words.
Address 116-117 (Flush Time) (Format: Time 1)
Get / Set the instrument's Flush Time. The Flush Time consists of 2 words.
Address 118 (Idle Mode) (Format: 16 Bit Word)
Get / Set the instrument's Idle Mode options. The Idle Mode consists of 1 word which
refers to 0 = Flow without Conductivity, 1 = Flow with Conductivity and 2 = No Flow.
Address 119 (Bypass flow) (Format: 16 Bit Word)
Get / Set the instrument's Bypass flow flag. The Bypass flow consists of 1 word which
refers to 0 = Disabled and 1 = Enabled.
Address 120 (Clean Time mode) (Format: 16 Bit Word)
Get / Set the instrument's Clean time mode flag. The Clean Time mode consists of 1
word which refers to 0 = Disabled and 1 = Enabled.